Home PrintPrint
Ì8-500                              Ì8-500U 


Oscilloscope-card

 


Application

Very high-frequency analog and digital data collection, quick processes and wide band signal capture.

Special features

  • 2 multiplexing analog inputs
  • input impedance: 1MOhm, 17pF and 50 +/-1 Ohm
  • input defence +/- 150 V
  • convertion time 2 ns
  • maximum sampling rate 500 MHz
  • PCI interface bus: Bus-Master
  • real dynamic range 53 dB


Application

Is widely used in science technology and industry

  • data capturing digital oscilloscope
  • quick processes registrator
  • panoramic wide band signal monitoring
  • non-destroying control
  • nuclear investigations

    Òechnical chracteristics

    Analog digital channel 
    Number of analog inputs 2 multiplexing
    Input impedance 1 ÌOhm, 17 pF (input 0)
    50 Îhm (input1)
    Bandwidth (-3 dB) 100 ÌÃö (channel 0)
    180 ÌÃö (channel 1)
    Input signal ranges(program setting) ±5V; ±2,5V; ±1V; ±0,5V; (channel 0)
    ±2,5V; ±1V;±0,5V; ±0,2V; (channel 1)
    Input channels voltage defence (power switched on) ±150Â (channel 0)
    ±15Â (channel 1)
    Memory buffer Ìàõ 2048 kbyte
    Memory buffer organization Shotly data collection with variable number of shots.The shot dimension the dimension of pre-history and history is programmed multiple power 2.
    Device-PC data exchange Program reading of the result of convertion in Slave mode after interruption or after setting the bit of readiness; DMA Bus-Masterchannel transfer. DMA transfer is distributed in memory (Scatter-Gatter).
    Analog-digital converter
    Resolution 8 bit
    Convertion time 2 ns
    Maximum sampling rate 500 ÌHz
    ADC start From internal quartz generator
    External clock frequency.
    External clock frequency Signal frequency upto 500MHz. The signal continues more than 1ns. Range: 0,3V-1V.
    Triggering
    Source Internal/External
    External trigger signal input voltage ranges (program setting) ±5V, ±1V
    Type Front or back
    Number of levels 256
    Requirements for triggering (program setting) Enter opened
    Enter closed
    VCH (range: 50kHz-100MHz)
    NCH (range: 50Hz-1MHz)
    General data
    PC interface bus PCI
    Power required +5V; 1,7À / +12V, 1,5À
    Shape 122õ220õ29 mm
    Weight 350 g

    Ì8-500 è Ì8-500U Cards test results

     oscilloscope program (~1,4 Ìb)